Development of a Versatile Mechanical Testing Device for In Situ Synchrotron Tomography and Diffraction Experiments - Matériaux (MAT) Accéder directement au contenu
Article Dans Une Revue Integrating Materials and Manufacturing Innovation Année : 2019

Development of a Versatile Mechanical Testing Device for In Situ Synchrotron Tomography and Diffraction Experiments

Maxime Pelerin
  • Fonction : Auteur
  • PersonId : 1057916
Andrew King
Lucien Laiarinandrasana
Henry Proudhon

Résumé

A new mechanical stage to perform in situ 3D imaging using synchrotron X-ray tomography is presented. Pairing control and acquisition allows the running of high quality continuous mechanical tests to study damage and fracture in any kind of structural materials. The modular design make this device very versatile with the possibility to use many specimen geometries and load ranges up to 5 kN, and switch within minutes from tomography to X-ray diffraction configurations. Examples of successful experiments to study the damage mechanisms with a technical polymer material are given
Fichier principal
Vignette du fichier
Pelerin_IMMI_2019.pdf (515.36 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-02355866 , version 1 (12-10-2021)

Licence

Paternité

Identifiants

Citer

Maxime Pelerin, Andrew King, Lucien Laiarinandrasana, Henry Proudhon. Development of a Versatile Mechanical Testing Device for In Situ Synchrotron Tomography and Diffraction Experiments. Integrating Materials and Manufacturing Innovation, 2019, 8 (3), pp.378-387. ⟨10.1007/s40192-019-00143-6⟩. ⟨hal-02355866⟩
52 Consultations
139 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More