Two-photon microscopy with simultaneous standard and extended depth of field using a tunable acoustic gradient-index lens. - École polytechnique Accéder directement au contenu
Article Dans Une Revue Optics Letters Année : 2009

Two-photon microscopy with simultaneous standard and extended depth of field using a tunable acoustic gradient-index lens.

Résumé

We describe a simple setup that allows depth of field switching at kilohertz rates in a nonlinear microscope. Beam profile and/or divergence are modulated using a tunable, acoustically driven gradient-index fluid lens. We demonstrate two modulation strategies, one based on fast varifocus scanning during each pixel and the other based on pseudo-Bessel beam excitation. Average beam shape is switched every line during scanning, resulting in the interlaced acquisition of two different images. We apply this approach to the simultaneous standard and 4.5x-extended depth-of-field imaging of developing embryos. © 2009 Optical Society of America
Fichier principal
Vignette du fichier
ol-34-11-1684.pdf (566.43 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-00818495 , version 1 (13-11-2013)

Identifiants

Citer

Nicolas Olivier, Alexandre Mermillod-Blondin, Craig B. Arnold, Emmanuel Beaurepaire. Two-photon microscopy with simultaneous standard and extended depth of field using a tunable acoustic gradient-index lens.. Optics Letters, 2009, 34 (11), pp.1684-1686. ⟨10.1364/OL.34.001684⟩. ⟨hal-00818495⟩